FujiFilm PROFECT CS Service Manual page 471

Table of Contents

Advertisement

10.
Once the initialization of the application is completed, press the [Enter]
key.
REFERENCE
The initialization of the application takes about 2 minutes.
11.
Turn OFF the power of the RU.
12.
Restore the DipSW1 No. 6 switch to its original setting.
13.
Reinstall the cover.
- Upper right-hand side cover
{MC:3.1_Covers}
14.
Restore the CL's IP address to the value recorded in the step 4 (user
specified value).
011-202-04
11.15.2004 FM4425
15.
Turn ON the power of the RU.
REFERENCES
- It takes about three minutes until the screen appears on the touch panel after
the RU is powered ON.
- When the RU is powered ON, the following screen appears on the touch panel,
and it boots up to the Maintenance Mode.
- When RU software version 1.2 or later is used, the [Return] button in the lower
right corner of the touch panel is hidden.
CR-IR 363 Service Manual
Appx MT-3
Appx MT-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Clearview csClearview csmCr-ir 363