Error
Error Name
Code
12842
Power supply B161 error
12843
Power supply B171 error
12844
Power supply B181 error
12845
Power supply B191 error
12846
Power supply B201 error
12847
Power supply B211 error
12848
Power supply B221 error
12849
Power supply B231 error
12850
Power supply B241 error
12851
Power supply B251 error
12852
Power supply B261 error
12853
Power supply B281 error
Power supply B series detection
12854
error (2)
011-202-11E
06.29.2012 FM6046
During bootup and during routine processing, it is detected that the "B161" fuse (F1) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B171" fuse (F2) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B181" fuse (F3) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B191" fuse (F4) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B201" fuse (F5) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B211" fuse (F6) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B221" fuse (F7) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B231" fuse (F8) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B241" fuse (F9) on the DRV17B board is blown.
During bootup and during routine processing, it is detected that the "B251" fuse (F10) on the DRV17B board is
blown.
During bootup and during routine processing, it is detected that the "B261" fuse (F11) on the DRV17B board is
blown.
During bootup and during routine processing, it is detected that the "B281" fuse (F12) on the DRV17B board is
blown.
During bootup and during routine processing, even though it is detected that the voltage of the DRV17B-related
+24V power supply is not outputted, the output is detected at either of the detection points of "B161" through
"B281".
Occurrence Condition
CR-IR 363 Service Manual
MT-49
Analysis
Flow
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
MT-49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Clearview csClearview csmCr-ir 363

Table of Contents