SCPI Command Reference
E4980A/AL commands
:CORRection:OPEN[:EXECute]
Syntax
:CORRection:OPEN[:EXECute]
Description
Executes OPEN correction based on all frequency points. (No Query)
Equivalent key
[Meas Setup] - CORRECTION - OPEN - MEAS OPEN
:CORRection:OPEN:STATe
Syntax
:CORRection:OPEN:STATe {ON|OFF|1|0}
:CORRection:OPEN:STATe?
Description
Enables OPEN correction.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - CORRECTION - OPEN - ON
[Meas Setup] - CORRECTION - OPEN - OFF
:CORRection:SHORt[:EXECute]
Syntax
:CORRection:SHORt[:EXECute]
Description
Executes SHORT correction based on all frequency points. (No Query)
Equivalent key
[Meas Setup] - CORRECTION - SHORT - MEAS SHORT
:CORRection:SHORt:STATe
Syntax
:CORRection:SHORt:STATe {ON|OFF|1|0}
:CORRection:SHORt:STATe?
Description
Enables SHORT correction.
Parameter
ON or 1
338
Description
Enables OPEN correction.
Disables OPEN correction.
Description
Enables SHORT correction