Secondary Connector Faceplate (Left) - Thermo Scientific HyPerforma G3Lite User Manual

Bioreactor controller
Table of Contents

Advertisement

Chapter 1
|
G3Lite overview
Vent filter
heater 1
Main power in
Power in vessel
heater
Ground stud
Thermo Scientific

1.5.1 Secondary connector faceplate (left)

Connectors on the left faceplate are shown in Figure 1.9 and described
below:
Figure 1.9. Secondary connector faceplate (left).
Vent filter heaters—These connectors provide a paired
connection for each of the two (2) vent filter heaters. Power is
supplied through the receptacles. Note: Because the voltage used
for the unit is carried through to the Vent Filter Heater receptacles,
you must ensure that the vent filter heaters are rated for the
appropriate voltage.
HMI/monitor—Supplies power for the computer monitor.
Power in/power out—These connectors are used to supply
power to the unit and is either 120 V AC or 240 V AC based on
facility requirements. Ensure that adequate power is supplied to the
Controller (15 A at 120 V AC or 6 A at 240 V AC).
HyPerforma G3Lite Bioreactor Controller User's Guide
HMI/monitor
Vent filter
heater 2
Power out
Agitator motor
Power out vessel
heater
Remote E-Stop
|
16

Advertisement

Table of Contents
loading