Installing The Hepa-Filter And The Water Reservoir Cover Panel; Figure 4-5. Assembling The Hepa-Filter And The Airbox - Thermo Scientific HERACELL VIOS 160i LK User Manual

Co2 incubator
Table of Contents

Advertisement

Installing the HEPA-Filter and the Water Reservoir Cover
Panel
The HEPA-filter
airbox rests on the water reservoir cover panel
1
2

Figure 4-5. Assembling the HEPA-Filter and the Airbox

1. Place the HEPA-filter
2. Tilt the airbox
corresponding slots provided in the HEPA-filter
3. Insert the tabs
(6/Figure
4. Place the seal
it down evenly.
Thermo Scientific
(2/Figure
4-5) must be inserted into the airbox
5
6
7
8
(2/Figure
4-5) on a flat surface.
(1/Figure
4-5) to the left and insert the tabs on the left
(5/Figure
4-5) on the right side of the airbox into the HEPA-filter catches
4-5).
(8/Figure
4-5) into the groove

Installing the HEPA-Filter and the Water Reservoir Cover Panel

(2/Figure
4-6) and is pushed against the air intake.
3
4
(4/Figure
(7/Figure
4-5) provided at the airbox tubing and press
HERA
(1/Figure
4-5) from below. The
4
(3/Figure
4-5) into the
4-5).
160i LK / 250i LK
CELL VIOS
4
Start-up
3
4-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heracell vios 250i lk

Table of Contents