Troubleshooting
Status Messages
6-8
Model 5030 SHARP Monitor Instruction Manual
Detailed Error Status
0000 0000 0000 00
A Sum status of the data and program memory (memory error)

Table 6-5. Data and Program Memory

Code
Definition
01
FLASH-EEPROM Error
02
RAM error
04
Save RAM error (backup battery exhausted)
08
EEPROM write/read error
10
free
20
free
40
free
80
free
0000 0000
00
B Sum status of the sampling and measuring system

Table 6-6. Sampling and Measuring System

Code
Definition
01
detector error < 1000 1/s ...> 500000 1/s
02
automatic HV adjustment not possible (<1200V ... > 1500V)
04
mass compensation >
08
pump off protection (do not block the air intake!)
10
lift sensor: "open" position not noticed
20
filter break
40
Plateau Flag (Counter
80
pump was switched off when nominal air flow could not be reached
00
00
±
10%
10)
(Table
6-5).
(Table
6-6).