Operation

Programmed Operation

HOLD
If you selected continuous operation (see
manually. Press the STOP key on the control panel. The centrifuge will be decelerated at the
designated rate. The message "END" will illuminate, and after pressing the key OPEN, the lid will
open and you can remove the samples.
Programmed Operation

Storing Selections to Program Memory

1. Establish all desired run parameters as explained above,
Note If you plan on using the QC RUN feature, consider the following as you save runs to
program memory:
• QC programs should all apply to using the same rotor.
• QC RUN programs should not use HOLD.
To minimize the time it takes to perform a QC RUN, we recommend the following:
• Starting with program 1, save runs to program memory: first by temperature (highest first), then,
by speed (lowest first).
FOR EXAMPLE: The programs below are listed in correct order for using QC RUN function
(program values shown are for example purposes only, actual set parameters will vary).
ROTOR
H-12000
H-12000
H-12000
H-12000
2. Press the MENU key to access the advanced features screens.
3. Press the ENTER key (six times after first accessing the advanced features) until the following
screen is in the SET DISPLAY:
4. Press 1.
The following screen will appear in the SET DISPLAY:
Note Single-digit program numbers 1-9 are for storing sets of standard run parameters;
double-digit program numbers 10-15 are for storing step-runs.
4-6
RC 12BP+
"HOLD"
SPEED
TIMED
2900
04:00
4000
03:00
3800
03:45
4700
05:30
on
page
4-4), you will have to stop the centrifuge
"Normal Operation"
DEG C
PROG #
22
1
22
2
4
3
4
4
on
page
4-3.