Calibration

German Networking Protocol

German Networking
Protocol
Storing Calibration
Values and
Operational
Parameters
4-12
Model 5030 SHARP Monitor Instruction Manual
Values From EEPROM in the next screen, all Operational and Calibration
changes will be lost. To store these settings, press YES.
All parameters, dates and calibration data are written into the EEPROM.
These data are reloaded, either in case an error is detected in the
battery-buffered RAM when switching the main switch or by simply giving
the respective command via keyboard entry in the next screen.
Depending upon the air monitoring program
requirements, the standard temperature can be
adjusted as needed using the +/- keys. Press
NEXT to continue.
The GESYTEC Protocol references network
protocols in Germany and can be set to either
I-N Compatible or to a STANDARD version.
Press SET to switch between version required by
the user.
Press NEXT to continue.
In the following two screens the number of
variables (1-7) and the value for each variable can
be chosen as necessary. Press NEXT.
Press NEXT.
This Calibration menu asks the user if they
would like to set the new parameters (such as,
Calibration and Operation) to be stored into the
EEPROM. If this is not done, the next time the
power is cycled, or a user chooses to Reload
Thermo Fisher Scientific