Troubleshooting
Status Messages
Overlay of Several Status
Messages
6-10
Model 5030 SHARP Monitor Instruction Manual
0000
00
00 0000 0000
F Sum status of the temperature measurement

Table 6-10. Temperature Measurement

Code
Definition
01
temperature T1 over range
02
temperature T2 over range
04
temperature T3 over range
08
temperature T4 over range
00
00
0000 0000 0000
G Sum status of nephelometer measurement

Table 6-11. Nephelometer Measurement

Code
Definition
01
nephelometer output over range
The error codes of the detailed error status messages are added and thus
form the detailed error status and the global error (failure) status
(Table
6-12).

Table 6-12. Detailed Error Status and Global Error Status

Error Status
01: 0000 0000 0000 0004
02: 0000 0000 0000 3000
04: 0000 0000 0001 0000
(Table
(Table
Description
SaveRAM error (backup battery empty)
Lifting position: opening of the filter tape gap has not
been recognized
Filter tape fissure: no filter tape transport has been
recognized
Differential pressure sensor
6-10).
6-11).