Status:operation Event Enable Register 2-1 - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Status Registers
Use Status Registers to Determine the State of Analyzer Events and Condition s
Bit
Description
0
A 1 in this bit position indicates that the analyzer i s
performing a self-calibration .
1, 2
Unused. These bits are always set to O .
3
A 1 in this bit position indicates that a sweep is i n
progress .
4$
A 1 in this bit position indicates that a measurement is in
progress .
5
A 1 in this bit position indicates that a measurement is in
a "wait for trigger" state .
6, 7
Unused . These bits are always set to O .
8b
A 1 in this bit position indicates that the instrument is i n
the paused state of the measurement.
9-14
Reserved. These bits are not used by the analyzer, but ar e
for future use with other Agilent products .
15
Always Zero (0) .
a. The description of this bit refers to any measurement unde r
the MEASURE key.
b. This bit applies to ESA optional measurement personalitie s
only, and may
or
may not be implemented in all such person-
alities.
STATus:OPERation Event Enable Registe r
The STATus:OPERation event enable register lets you choose the bit s
that will set the operation status summary bit (bit 7) of the status byt e
register to 1 . Send the : STATus : OPERation : ENABle <u m> command
where <num> is the sum of the decimal values of the bits you want to
enable .
For example, to enable bit 9 and bit 3 (so that whenever either bit 9 or 3
is set to 1, the operation status summary bit of the status byte registe r
will be set to 1), send the : STATus :OPERation : ENABle 520 (512 + 8 )
command. The : STATus : OPERation :ENABle? command returns the
decimal value of the sum of the bits previously enabled with the
: STATus : OPERation : ENABle <num> command.