Language Reference
[SENSeJ :MIXer Subsectio n
Set External Mixer Bias On/Off
[ :SENSe] :Mlxer :BIAS :STATe <boolean >
[ :SENSe] :MIXer :BIAS :STATe ?
This command activates an internal bias source for use with externa l
mixers . This command is available only on model E4407B analyzers
having Option AYZ .
Factory Prese t
and *RST:
Of f
Remarks:
The bias signal is present on the center conductor of th e
IF Input connector on the front panel . See related
command [ : SENSe] :MIXer :BIAS <numeric> .
NOTE
Mixer Bias will be Off if AUTO Harmonic and Presel Mixer Type i s
selected .
Front Panel
Access:
Input/Output (or Input), Input Mixer, Mixer Confgg, Mixer
Bias On Off
Set External Mixer LO Harmonic Value
[ :SENSe] :MIXer :HARMonic <value >
[ :SENSe] :MIXer :HARMonic ?
This command allows you to set the LO harmonic value for mixers
other than the HP/Agilent 11970-Series or 11974-Series Mixers . This i s
done after the mixer harmonic configuration
([ : SENSe] :Mixer : HARMonic :AUTO <boolean>) is set to manual .
This command is available only on model E4407B analyzers having
Option AYZ.
Factory Prese t
and *RST:
-8 (Band A, 26 .5-40 GHz )
Range :
Any non-zero integer from -50 to 50, inclusiv e
Remarks:
The harmonic value with its associated sign i s
automatically determined from the external mixe r
band selected . This is the AUTO mode . For mixers
other than the HP/Agilent 11970-Series o r
11974-Series, an LO harmonic other than that
determined in the AUTO mode may be required . This i s
achieved using this command and related comman d
[ : SENSe ] : MI Xer : HARMonic : AUTO <boolean> .