Language Reference
[ :SENSe] :AVERage Subsection
[ :SENSe] :AVERage Subsection
Clear the Current Average
[ :SENSe] :AVERage :CLEar
Re-start the trace averaging function .
NOTE
Re-start the trace at the beginning of a sweep to obtain valid averag e
data. To do this, remotely abort the sweep and initiate a single sweep .
Front Panel
Access:
none
Set the Average Count
[ :SENSe] :AVERage :COUNt <integer>
[ :SENSe] :AVERage :COUNt ?
Specifies the number of measurements that are combined.
Factory Preset
and *RST :
10 0
Range :
1 to 819 2
Front Panel
Access:
BW/Avg, Average On Of f
Turn Averaging On/Off
[ :SENSe] :AVERage[ :STATe] OFFIONIOI 1
[ : SENSe] :AVERage[ :STATe] ?
This command toggles averaging off and on . Averaging combines the
value of successive measurements to average out measuremen t
variations .
Factory Preset
and *RST:
Of f
Remarks:
When a measurement under the front panel
MEASUR E
key is started, this command is turned off for video
averaging ([ : SENSe] :AVERage : TYPE LPOWer) . If thi s
command is turned on for video averaging when any of
the
MEASURE
key measurements are in progress, that
measurement will be stopped.
Front Panel
Access :
BW/Avg, Average On Off
Chapter 5
5-85