Error Message s
Greater than 1000 : Personality Specific Error Message s
Greater than 1000 :
Personality Specific Error Message s
An error detected with a number greater than 1000 indicates the
instrument has detected an error relating to an installed personality .
For more information on these errors, refer to the user's guide for the
personality in use.
10219
Awaiting trigger
A trigger event from the selected trigger source has not
been detected.
10164
Band measurement not defined for Out of Band .
User is attempting to monitor the band but has set the
frequency outside the band . Reset the band for th e
particular standard for which you are testing or use the
channel setting which does not require a frequency t o
be set . (Meas Setup, Method (Channel)) .
10228
Cannot correlate to input signal .
This error is normally generated because of one of th e
following reasons: 1 . There is no carrier signal .
2. Walsh channels other than the pilot are active.
3. There is some other modulation problem that wil l
prevent the measurement from being made .
This problem must be corrected before th e
measurement can continue .
10163
Cannot find the Power vs Time Limits File .
The limit line definition file for the GSM standards ha s
been deleted. This message is displayed while the
Measure key is greyed out. Reinstall the GSM
measurement personality.
10166
Cannot update the list of cable types .
The cable file may have been moved or deleted
accidentally. Reinstall the GSM measurement
personality.