Agilent Technologies ESA-E Series Programmer's Manual page 289

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
SOURce Subsyste m
Set the Output Power Step Size
SOURce :POWer :STEP[ :INCRement] <ampl>
:SOURce :POWer :STEP[ :INCRement] ?
Specifies the source power step size .
Default Unit: dB m
Front Panel
Access:
Source, Amptd Step Auto Ma n
Set the Source Sweep Power Range
SOURce :POWer :SWEep <rel_ampl>
SOURce :POWer :SWEep ?
Specifies the range of power levels through which the source output wil l
sweep. Use : SOURce : POWer : STARt to set the power level at the star t
of the power sweep . See also : SOURce : POWer : SPAN.
Factory Prese t
and *RST:
0 dB
Range :
0 dB to 20 dB
Default Unit: dB
Front Panel
Access :
Source, Power Sweep On Of f
Output Power Tracking
SOURce :POWer :TRCKing <number>
SOURce :POWer :TRCKing ?
Adjusts the tracking of the source output with the spectrum analyze r
sweep in the present resolution bandwidth.
Factory Preset
and *RST:
This command is persistent . The term persistent
means that the command retains the setting previousl y
selected, even through a power cycle .
Range:
Real, 0 to 4095
Remarks:
This command is not needed with the 1 .5 GHz trackin g
generator .
Front Panel
Access:
Source, Man Track Ad j
Chapter 5
5-129

Advertisement

Table of Contents
loading

Table of Contents