Agilent Technologies ESA-E Series Programmer's Manual page 362

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Error Messages
201 to 799 : Device-Specific Errors
223
Trigger Offset unavailable in swept span s
Trigger Offset is only available in Zero Span . Refer to
Agilent Technologies ESA Spectrum Analyzers User' s
Guide for a description of this function .
215
TG start freq is less than 1/2 res bw
TG uncalibrated at start frequencies below 1/2 th e
current resolution bandwidth .
214
TG start freq is less than 9kH z
TG uncalibrated below 9 kHz .
204
TG Frequency Limit
The tracking generator has reached the limit of it s
allowable frequency range .
736
Too
many data values at <freq_or_time value >
This message may appear when data is sent to a
corrections or limit table using the DATA or
MERG E
commands. These tables limit the number of
amplitudes associated with a frequency or time to 2 o r
less. This message will appear if an attempt is made t o
attach 3 or more values to a frequency or time .
206
Unable to initialize flatness dat a
A failure occurred in setting the flatness data in the
internal EEROM. Contact your local Agilent
Technologies sales and service office .
762
Unable to load file
A failure occurred while loading a file; the file was not
loaded.
759
Unable to load state saved from firmwar e
Rev A .03 .00
A saved state file from a newer firmware revision wa s
attempted to be loaded into an older instrument .
7-34
Chapter 7