Error Messages
201 to 799 : Device-Specific Errors
201 to 799:
Device-Specific Errors
Some device operations did not properly complete, possibly due to an
abnormal hardware or firmware condition . These codes are also used
for self-test response errors . Errors in this class set the device-specifi c
error bit (bit 3) in the event status register (IEEE 488 .2, section 11 .5.1).
The <error_message> string for
a positive
error is not defined by SCPI .
A positive error indicates that the instrument detected an error withi n
the GPIB system, within the instrument firmware or hardware, during
the transfer of block data, or during calibration .
653
Auto Align not available when usin g
Calibration Default s
The Auto Alignment system cannot be used until a n
Align Now All is
executed by pressing
System ,
Alignments, Align Now, All .
For Agilent Technologie s
E4402B, E4403B, E4404B, E4405B, E4407B, an d
E4408B only, you must connect the
AMPTD REF OUT
to
-rr--r
r
alignment.
614
Bad or missing floppy dis k
The floppy is not inserted or the directory could not b e
read. Insert a known good disk and try again .
205
Command not recognize d
Indicates that the command sent from the remote
interface was not recognized . Check the programming
guide for correct syntax.
205
Command not recognized
Indicates that the command sent from the remot e
interface was not recognized. Check the programmin g
guide for correct syntax .
Chapter 7
7-27