Agilent Technologies ESA-E Series Programmer's Manual page 214

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
FORMat Subsyste m
NOTE
This command specifies the formats used for trace data during data
transfer across any remote port .
For corrected trace data ( : TRACe [ : DATA] with paramete r
<trace_name>), REAL and ASCii formats will provide trace data i n
the current amplitude units . INTeger format will provide trace data i n
mdBm. The fastest mode is INTeger, 32 .
For uncorrected trace data ( : TRACe [ : DATA ] with parameter
RAWTRACE), UINTeger and INTeger formats apply to RAWTRACE
queries, and return uncorrected ADC values. The fastest mode is
UINTeger ,16 .
For state data, the format cannot be changed. It is always in a machine
readable format only (machine units) .
Table 5- 1
Corrected Trace Data Types
:TRACe :DATA?<trace name >
Data Type
Result
ASCII
Amplitude Units
INT,32
(fastest)
Internal Units
REAL,32
Amplitude Unit s
REAL,64
Amplitude Units
Table 5-2
Uncorrected Trace Data Type s
:TRACe :DATA? RAWTRAC E
Data Type
Result
INT,32
Uncorrected
AD C
Value s
UINT,16
(fastest)
Uncorrected
ADC
Value s
ASCii -
Amplitude values are in ASCII, in amplitude units ,
separated by commas .
INTeger,
32 - Binary 32-bit integer values in internal units (mdBm) ,
in a definite length block.
REAL, 32 (or 64) - Binary 32-bit, or 64-bit, real values in amplitud e
units), in a definite length block.
5-54
Chapter5