Status:operation Event Enable Register - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Use Status Registers to Determine the State of Analyzer Events and Conditions
Bit
0
A 1 in this bit position indicates that the analyzer is
performing a self-calibration.
1, 2
Unused. These bits are always set to 0.
3
A 1 in this bit position indicates that a sweep is in
progress.
A 1 in this bit position indicates that a measurement is in
a
4
progress.
5
A 1 in this bit position indicates that a measurement is in
a "wait for trigger" state.
6, 7
Unused. These bits are always set to 0.
b
A 1 in this bit position indicates that the instrument is in
8
the paused state of the measurement.
9–14
Reserved. These bits are not used by the analyzer, but are
for future use with other Agilent products.
15
Always Zero (0).
a. The description of this bit refers to any measurement under
the
key.
MEASURE
b. This bit applies to ESA optional measurement personalities
only, and may or may not be implemented in all such person-
alities.

STATus:OPERation Event Enable Register

The STATus:OPERation event enable register lets you choose the bits
that will set the operation status summary bit (bit 7) of the status byte
register to 1. Send the :STATus:OPERation:ENABle <num> command
where <num> is the sum of the decimal values of the bits you want to
enable.
For example, to enable bit 9 and bit 3 (so that whenever either bit 9 or 3
is set to 1, the operation status summary bit of the status byte register
will be set to 1), send the :STATus:OPERation:ENABle 520 (512 + 8)
command. The :STATus:OPERation:ENABle? command returns the
decimal value of the sum of the bits previously enabled with the
:STATus:OPERation:ENABle <num> command.