Language Reference
MEASure Subsystem
Measure Upper Channel Powe r
:MEASure :ACPower :UPPer ?
This command returns the value of the upper channel power relative t o
the main channel power, in dB.
Measure Channel Power and Densit y
:MEASure :CHPower ?
This command returns scalar results of main channel power, and powe r
density.
Remarks:
The main channel power is returned in the curren t
amplitude units, and the density value is returned i n
current amplitude units/Hz . The results are returned
in a comma-separated list.
Measure Channel Powe r
:MEASure :CHPower :CHPower ?
This command returns the value of the channel power in amplitud e
units.
Front Panel
Access :
MEASURE, Channel Power
Measure Channel Power Density
:MEASure :CHPower :DENSity ?
This command returns the value of the channel power density in
amplitude units/Hz .
Measure Emission Bandwidt h
:MEASure :EBWidth ?
:MEASure :EBWidth :EBWidth ?
This command returns the value of emission bandwidth in Hz .
Front Panel
Access :
MEASURE, Emission BW
Return Harmonic Amplitude s
:MEASure :HARMonics :AMPLitude :ALL ?
Returns a comma-separated list of the amplitudes of the ten harmonics .