Error Message s
201 to 799: Device-Specific Errors
769
Invalid instrument mode
You have attempted to switch to an instrument mod e
that is currently not installed. Confirm that the mod e
name (for
INST : SEL)
or number (for
INST : NSEL)
wa s
entered correctly and that the requested personality is
actually installed in the instrument .
221
Invalid option, unable to uninstall packag e
You have attempted to remove a personality that is no t
currently installed . Verify command was entered
correctly.
701
Invalid printer response
In attempting to identify the printer an invalid
response was received . Check that you are using a
supported printer . Be sure you are using the prope r
cable and that it is securely fastened .
606
Media is corrupt
A save was attempted to a corrupt device.
609
Media is not writabl e
A save was attempted to a read-only device .
605
Media is protected
A save was attempted to a write-protected device .
202
No
peak found
No signal peak was found.
201
Option not installe d
The desired operation cannot be performed because a
required option is not installed . For example, pressing
Source with no tracking generator installed in the
analyzer will generate this error.
7-32
Chapter 7