Agilent Technologies ESA-E Series Programmer's Manual page 186

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Referenc e
CALCulate :MARKer Subsection
Sets the resolution of the marker frequency counter so it i s
automatically coupled to the frequency span, generating the fastes t
accurate count.
Factory Prese t
and *RST :
On
Front Panel
Access :
Freq Count, Resolution Auto Man
Frequency Counter Marke r
:CALCulate :MARKer[1]121314 :FCOunt[ :STATe] OFFIONIO1 1
:CALCulate :MARKer[1]121314 :FCOunt[ :STATe] ?
:CALCulate :MARKer[1] :FCOunt :B?
Turns on or off the marker frequency counter. To query the frequency
counter, use :CALCulate :MARKer[1] :FCOunt :X?
Factory Preset
and *RST :
Of f
Remarks:
If query with frequency count off, 9e15 is returned .
Front Panel
Access :
Freq Count, Marker Count On Of f
Marker Function
:CALCulate :MARKer[1]121314 :FUNCtion BPOWerINOISeIOFF
:CALCulate :MARKer[1]121314 :FUNCtion ?
Selects the marker function for the specified marker. To query the value
returned by the function, use : CALCulate : MARKer [ 1 ] ~ 2 ~ 3 ~ 4 :Y?
BPOWer is the power integrated within the bandwidt h
NOISe is a noise measurement
OFF turns off all functions
Remarks:
When a measurement under the front panel MEASURE
key is started, this command is turned off. If this
command is turned on when any of the MEASURE key
measurements are in progress, that measurement will
be stopped.
Front Panel
Access :
Marker, Function