Language Reference
[ :SENSe] :DEMod Subsectio n
Demod Time
[ :SENSe] :DEMod :TIME <time>
[ :SENSe] :DEMod :TIME ?
Sets the time used for frequency domain demodulation .
Factory Preset
and *RST:
500 ms
Range :
2 ms to 100 s
Default Unit: second s
Front Panel
Access :
Det/Demod, Demod, Demod Tim e
Demod Vie w
[ :SENSe] :DEMod :VIEW[ :STATe] OFFI ONI 01 1
[ :SENSe] :DEMod :VIEW[ :STATe] ?
This command causes the demodulated signal to be displayed . If FM
Demod is on, then the display scales the y-axis in units of kHz. The
scale/div is set with the command : DISP1ay : WINDow: TRACe : Y
[ :SCALe] :PDIVision :FREQuency <freq> if FM Demod is on . If FM
Demod is on, then several functions are not available ; these include:
Log/Lin (display is always in linear), Y -Axis Units, Marker Searc h
functions, Normalize, Display Line, Peak Excursion, and Pea k
Threshold. There is no effect when AM demodulation is used (onl y
applicable for FM demodulation) .
Factory Preset
and *RST:
Of f
Remarks:
This command is not available when Demod is set to
Off.
Front Panel
Access :
Det/Demod, Demod, FM, Demod Vie w