Language Reference
STATus Subsystem
Operation Negative Transitio n
STATus :OPERation :NTRansition <number>I<non-decima l
numeric >
STATus :OPERation :NTRansition ?
This command determines which bits in the Operation Conditio n
register will set the corresponding bit in the Operation Event registe r
when that bit has a negative transition (1 to 0) . The variable <number >
is the sum of the decimal values of the bits that you want to enable.
Factory Prese t
and *RST:
0
Range :
0 to 3276 7
Operation Positive Transitio n
STATus :OPERation :PTRansition <number>I<non-decima l
numeric>
STATus :OPERation :PTRansition ?
This command determines which bits in the Operation Conditio n
register will set the corresponding bit in the Operation Event registe r
when that bit has a positive transition (0 to 1) . The variable <number >
is the sum of the decimal values of the bits that you want to enable .
Factory Prese t
and *RST:
32767 (all 1's)
Range :
0 to 32767
Preset the Status Byte
:
STATus : PRESet
Sets bits in most of the enable and transition registers to their defaul t
state. It presets all the Transition Filters, Enable Registers, and th e
Error/Event Queue Enable. It has no effect on Event Registers ,
Error/Event Queue ESE, and SRE Registers as described in IEE E
Standard 488 .2-1992, IEEE Standard Codes, Formats, Protocols and
Common Commands for Use with ANSI /IEEE Std 488 .1-1987 .
New York, NY, 1992 .