Language Reference
[ :SENSe] :SWEep Subsectio n
Factory Preset
Full Span is activated
Last Span is activated
On entering a measurement
( :MEASure
subsystem
commands )
While segmented sweep is on, the following function s
are unavailable :
Auto couple
BW/Avg RBW, VBW
Center/Start/Stop frequency, CF Step, an d
frequency Offset
Span, sweep time, and point s
Front Pane l
Access :
Sweep, Segmented, ON/OF F
Sweep Time
[ :SENSe] :SWEep :TIME <time>
[ :SENSe] :SWEep :TIME ?
Specifies the time in which the instrument sweeps the display .
Factory Prese t
and *RST:
ESA E4401B, E4411B : 4ms
ESA E4402B, E4403B : 5 ms
ESA E4404B : 16.75 ms
ESA E4405B : 33 ms
ESA E4407B, E4408B : 265 ms
Range:
The range depends upon the installed options, number
of sweep points, and firmware revision of your
instrument . See
"Sweep Time Range"
in the
Specifications Guide for details .
Default Unit: seconds
Remarks:
A span value of 0 Hz causes the analyzer to enter zer o
span mode. In zero span the X-axis represents time
rather than frequency. In this mode, the sweep time
may be set to faster values when Options AYX or B7 D
are installed.
Front Panel
Access :
Sweep, Sweep Time Auto Ma n
5-122
Chapter 5