Language Reference
CALCulate:LLINe Subsection
Test the Data Against the Limit Lin e
:CALCulate :LLINe[1]12 :FAIL?
Queries the status of the limit line testing . Returns a 0 if the dat a
passes, and returns a 1 if there is a failure . This query only make s
sense if margin or limit test is On. Use the command
: CALCulate : LLINe [1 ] ~ 2 : STATe
OFF ~ ON ~ 0 ~ 1
to activate limit line
testing.
Set the Margin Size
:CALCulate :LLINe[1]12 :MARGin <ampl_rel >
:CALCulate :LLINe[1]I2 :MARGin ?
Allows you to define the amount of measurement margin that is adde d
to the designated limit line.
Factory Prese t
and *RST :
not affecte d
Default Units : dB
Remarks:
The margin must be negative for upper limit lines, an d
positive for lower limits.
r
FOUL
.rUlle l
Access :
Display, Limits, Modify, Margin On Of f
Display the Limit Margin
:CALCulate :LLINe[1]12 :MARGin :STATe
OFFIONIO1 1
:CALCulate :LLINe[1]12 :MARGin :STATe ?
Allows you to display a measurement margin that is added to th e
designated limit line to do secondary testing of the data .
Factory Preset
and *RST:
Of f
Front Panel
Access :
Display, Limits, Modify, Margin On Off
Control Limit Line Testing
:CALCulate :LLINe[1]12 :STATe
OFFIONIO1 1
:CALCulate :LLINe[1]I2 :STATe?
Chapter 5
5-23