Standard Event Status Register 2-1 - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Status Registers
Use Status Registers to Determine the State of Analyzer Events and Condition s
Standard Event Status Registe r
The standard event status register is used to determine the specifi c
event that sets bit 5 in the status byte register. The standard event
status register does not have negative and positive transition registers,
nor a condition register. Use the IEEE common commands at th e
beginning of Chapter 5, "Language Reference" in this guide to acces s
the register.
To query the standard event status register, send the
*ESR
command .
The response will be the decimal sum of the bits which are set to 1. For
example, if bit number 7 and bit number 3 are set to 1, the decimal su m
of the 2 bits is 128 plus 8 . So the decimal value 136 is returned .
Figure 2-7
Standard Event Status Register
♦ To Status Byte Register Bit #5
ck723a
Operation Complete
Request Bus Control
Query Error
Device Dependent Erro r
Execution Erro r
Command Error
User Request
Power On
V
7
Event Register
7 6 5 4 3 2 1
0
Event
Enable Register
7 6 5 4 3 2 1
0
2-14
Chapter2