Agilent Technologies ESA-E Series Programmer's Manual page 181

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
CALCulate :LLUNe Subsection
Remarks :
If two amplitude values are entered for the same
frequency, a single vertical line is the result . In thi s
case, if an upper line is chosen, the amplitude of lesse r
frequency (amplitude 1) is tested . If a lower line is
chosen, the amplitude of greater frequency (amplitude
2) is tested.
For linear amplitude interpolation and linear frequency
interpolation, the interpolation is computed as:
Yi+1-Yi
(f-f 1
• )
+ Y i
Y =
f
i+1
-f
i
For linear amplitude interpolation and log frequenc y
interpolation, the interpolation is computed as :
y
Yi
+
1- Y i
= log fi + - log f;
(log f -log f i) + yi
1
For log amplitude interpolation and linear frequency
interpolation, the interpolation is computed as :
log y
i +1 -
log yi
log y =
(f - fi ) + log y i
f
i+1-
f
i
For log amplitude interpolation and log frequency
interpolation, the interpolation is computed as :
log
logyi+1-logYl(lo f-lo f)+to
y
= logfi+l- logfi
g
g
1
g
Yi
Front Panel
Access:
Display, Limits, Properties, X Axis Units Freq Tim e
Display, Limits, Modify, Edit
Display, Limits, Modify, Edit, Poin t
Display, Limits, Modify, Edit, Frequency
Display, Limits, Modify, Edit, Amplitude
Display, Limits, Modify, Edit, Connecte d
Display, Limits, Modify, Edit, Delete Point
Chapter 5
5-21

Advertisement

Table of Contents
loading

Table of Contents