Agilent Technologies ESA-E Series Programmer's Manual page 60

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Status Registers
Use Status Registers to Determine the State of Analyzer Events and Condition s
STATus :QLTEStionable:INTegrity:UNCalibrate d
Condition Register Bits
Bit
Decimal
Value
Descriptio n
0
0
Oversweep (Meas Uncal) A 1 in this position
indicates that the anlyzer is in a state that could lea d
to uncalibrated measurements. This is typically
caused by sweeping too fast for the curren t
combination of span, resolution bandwidth, and video
bandwidth. Auto coupling may resolve this problem .
1
2
Signal Ident ON A 1 in this bit position indicate s
that amplitude measurements may be in error due t o
signal identification routines being active . Amplitud e
accuracy is degraded when signal identification i s
active . (Agilent ESA model E4407B Option AYZ ,
External Mixing, only) .
2-14
-
Reserved These bits are not used by the analyzer, bu t
are for future use with other Agilent products .
15
32768
Always zero (0) This bit is always set to 0 .
2-26
Chapter 2