Language Reference
COUPle Subsyste m
COUPle Subsyste m
Some measurement settings are automatically coupled together t o
optimize speed and accuracy. These commands control that coupling.
COUPle the Function to Other Settings
:COUPle ALL NONE
:COUPle ?
The instrument can automatically couple instrument settings togethe r
for accurate measurements and optimum dynamic range . This
command is used to override the coupling for special measuremen t
needs.
Factory Preset
and *RST:
All
Remarks:
The following list of analyzer functions can either be
automatically coupled, or not coupled (manually set) :
Resolutionl bandwidt h
Video bandwidth
Sweep time
Center frequency step
Attenuatio n
Marker count resolutio n
Source attenuation
Source amplitude step
COUPle NONE
puts these functions into the manually
set (not coupled) mode .
COUPle ALL
puts the function s
into the auto coupled mode, and also puts the swee p
coupling mode into SA (couple all) .
Front Panel
Access :
Auto Couple
5-40
Chapter 5