Agilent Technologies ESA-E Series Programmer's Manual page 59

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Status Registers
Use Status Registers to Determine the State of Analyzer Events and Condition s
Bit
Decima l
Value
Description
4
16
IF/ADC Over Range The signal input level is to o
high, causing the analyzer analog-to-digital converte r
(ADC) range to be exceeded. This may occur wit h
resolution bandwidths less than or equal to 300 Hz in
zero span. (Agilent ESA models E4401B, E4402B ,
E4404B, E4405B, and E4407B only) .
5a
32
Over Range A 1 in this bit position indicates that the
signal is too large at the analog-to-digital converter
(ADC) .
6a
64
Under Range A 1 in this bit position indicates that
the signal is too small at the analog-to-digita l
converter (ADC).
7a
128
Insufficient Data A 1 in this bit position
indicates that there is not enough information t o
perform the measurement or function .
8a
256
Acquisition Failure A 1 in this bit positio n
indicates that the demod algorithm cannot correlat e
to the signal.
.,a
!19
M ni rv Prnh1 am A 1 in this hit nu-mil-inn inrlinata Q
Tailure
o2
the rile system memory or digital signal
processor (DSP) memory.
10 a
1024
Auto-Trigger Timeout A 1 in this bit position
indicates that the measurement timed out due to n o
trigger.
l la
2048
Trigger Problem A 1 in this bit position indicate s
that the measurement timed out due to no trigger.
12
4096
Invalid Data A 1 in this bit position indicates that
the present trace data does not reflect the existing
analyzer state. Trigger a new sweep and/or
measurement .
13a
8192
Unidentified Error A 1 in this bit position
indicates that a measurement has terminated for a
reason other than that given in any of the other bits .
14a
16384
Setting Limited/Readjusted A 1 in this bit
position indicates that the user settings could not be
achieved with the existing hardware ; values were set
to limits .
15
32768
Always Zero (0) This bit is always set to 0 .
a. This bit applies to ESA optional measurement personalitie s
only, and may or may not be implemented in all such person-
alities .
Chapter 2
2-25

Advertisement

Table of Contents
loading

Table of Contents