Agilent Technologies ESA-E Series Programmer's Manual page 204

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
DISPlay Subsystem
Control the Display Line
:DISPlay :WINDow:TRACe :Y :DLINe :STATe OFFIONIO1 1
:DISPlay :WINDow:TRACe :Y :DLINe :STATe ?
Turns the display line on or off .
Factory Preset
and *RST:
Of f
Front Panel
Access:
Display, Display Line On Off
Normalized Reference Level
:DISPlay :WINDow:TRACe :Y[ :SCALe] :NRLevel <rel ampl>
:DISPlay :WINDow:TRACe :Y[ :SCALe] :NRLevel ?
Sets the normalized reference level .
NOTE
See command : CALCulate : NTData [ STATe ] OFF
I
ON
I
01 1
Factory Preset
and *RST :
0 d B
Range :
-327 .6 to 327.6 dB
Default Unit: Current active units
Front Pane l
Access :
View/Trace, Normalize, Norm Ref Lvl
Normalized Reference Level Positio n
:DISPlay :WINDow:TRACe :Y[ :SCALe] :NRPosition <rel_ampl >
:DISPlay :WINDow:TRACe :Y[ :SCALe] :NRPosition ?
Selects the position of the normalized reference level. The top and
bottom graticule lines correspond to 10 and 0, respectively.
NOTE
See command : CALCulate : NTData [ STATe ] OFF
I
ON
I
01 1
Factory Prese t
and *RST:
10
Range :
0 to 10
Front Panel
Access :
View/Trace, Normalize, Norm Ref Posn

Advertisement