Language Referenc e
[ :SENSe] :CORRection Subsection
Amplitude correction is applied to the display data to adjust fo r
measurement situations where the unit under test has a differen t
impedance than the 5052 input impedance of the analyzer . Some Agilent
ESA analyzers have Option 1DP, 75Q input. In this case, you may want
to convert the data to make measurements in a 5052 system .
Factory Preset
and *RST:
The factory default is the input impedance of th e
analyzer.
Range :
50 or 75 ohms
Default Unit : ohms
Front Panel
Access :
Input, Input Z Corr 50 52 75 Q
External Amplifier Correction
[ :SENSe] :CORRection :OFFSet[ :MAGNitude ]
<rel_ampl >
[ :SENSe] :CORRection :OFFSet[ :MAGNitude] ?
A single value of amplitude correction can be applied to the displaye d
trace data to compensate for signal losses or gains that are due to other
devices in the measurement setup, rather than the unit under test .
Factory Preset
and *RST:
0 dB
Range :
-81 .9 to 81 . 9
Default Unit: dB
Front Pane l
Access :
AMPLITUDE/Y Scale, Ext Amp Gain
5-94
Chapter 5