Programming Examples
Using Limit Lines
/*
CAL :SOUR :STAT OFF
/* - The limit line test should fail .
/* -
Close the session .
/************************************************************ /
#include <stdio .h>
#include <stdlib .h>
#include <math .h>
#include <conio .h>
#include <ctype .h>
#include <string .h>
#include <windows .h>
#include "visa .h "
#define YIELD Sleep(5000 )
#define hpESA IDN_E4401B "Hewlett-Packard, E4401B "
#define hpESA IDN E4411B "Hewlett-Packard, E4411B "
ViSession defaultRM, viESA ;
ViStatus errStatus ;
ViChar
cIdBuff[256]= {0} ;
char
cEnter = 0 ;
int
iResult = 0 ;
long
lLimitTest =0L ;
/*Set the input port to 50MHz amplitude reference* /
void Route50MHzSignal( )
{
viQueryf(viESA, "*IDN?\n", "%t", &cldBuff) ;
iResult = (strncmp( cIdBuff, hpESA IDN_E4401B, strlen(hpESA IDN E4401B)) &&
strncmp( cIdBuff, hpESA IDN E4411B, strlen(hpESA IDN E4411B))) ;
if( iResult = 0 )
{
/*Set the input port to the 50MHz amplitude reference for the models* /
/*E4411B and E4401B* /
viPrintf(viESA,"CAL :SOUR :STAT ON \n") ;
/* For the analyzers having frequency limits >= 3GHz, prompt the user* /
/* to connect the amplitude reference output to the input* /
printf ( "Connect AMPTD REF OUT to the INPUT \n") ;
printf ("
Press Return to continue \n") ;
scanf( "%c",&cEnter) ;