Language Reference
CALCulate:MARKer Subsection
Marker Peak (Maximum) Searc h
CALCulate :MARKer[11 2~3I4 :MAXimum
Performs a peak search based on the search mode settings o f
CALCulate :MARKer :PEAK :SEARch :MODE .
NOTE
See command : CALCulate :MARKer : PEAK : SEARch :MODE
Front Panel
Access :
Peak Search (or Search), Meas Tools, Peak Searc h
Marker Peak (Maximum) Left Search
:CALCulate :MARKer[1]121314 :MAXimum :LEFT
Places the selected marker on the next highest signal peak to the left o f
the current marked peak .
Remarks :
The marker will be placed at the next highest peak that
rises and falls by at least the peak excursion above th e
peak threshold . If no peak meets the excursion and
threshold criteria, a No Peak Found error (202) is given .
Front Panel
Access :
Peak Search (or Search), Next Pk Left
Marker Next Peak (Maximum) Searc h
:CALCulate :MARKer[l]121314 :MAXimum :NEXT
Places the selected marker on the next highest signal peak from th e
current marked peak .
Remarks :
The marker will be placed at the highest peak that
rises and falls by at least the peak excursion above th e
peak threshold. If no peak meets the excursion an d
threshold criteria, a No Peak Found error (202) is given .
Front Panel
Access :
Peak Search (or Search), Next Peak
Marker Peak (Maximum) Right Search
:CALCulate :MARKer[l]121314 :MAXimum :RIGHt
Places the selected marker on the next highest signal peak to the right
of the current marked peak .
Remarks :
The marker will be placed at the highest peak tha t
rises and falls by at least the peak excursion above th e
peak threshold . If no peak meets the excursion and
threshold criteria, a No Peak Found error (202) is given .