Agilent Technologies ESA-E Series Programmer's Manual page 194

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
CALCuIate:MARKer Subsection
Position the designated marker on its assigned trace at the specified X
position . A trace is composed of 8192 points (X positions) when i n
non-zero span.
The query returns the current X position for the designated marker .
Range :
0 to 8191, (2 to 8192 in zero span for analyzers with
firmware revision A .05 .00 and later )
Front Panel
Access :
Marke r
Span Markers Center Frequency X Position
:CALCulate :MARKer[1]I21314 :X :POSition :CENTer <param>
:CALCulate :MARKer[1]I21314 :X :POSition :CENTer ?
Position the center frequency, of the designated span-type marker pair,
at the specified trace X position . A trace is composed of 8192 points (X
positions) when in non-zero span. Use : CALCulate : MARKer :MOD E
SPAN to select span markers .
The query returns the current X-position center frequency of th e
designated markers .
Range :
0 to 8191, (2 to 8192 in zero span for analyzers wit h
firmware revision A.05.00 and later)
Front Panel
Access :
Marker, <active marker>, Center
Span Markers Span X Position
:CALCulate :MARKer[1]I21314 :X :POSition :SPAN <param>
:CALCulate :MARKer[1]I21314 :X:POSition :SPAN?
Change the frequency span, of the designated span-type marker pair, t o
position the markers at the desired trace X positions . A trace is
composed of 8192 points (X positions) when in non-zero span . Use
: CALCulate :MARKer :MODE SPAN to select span markers .
The query returns the current X-position frequency span of th e
designated markers .
Range :
0 to 8191, (2 to 8192 in zero span for analyzers wit h
firmware revision A.05.00 and later)
Front Panel
Access :
Marker, <active marker>, Spa n
Band Markers Start Frequency X Position
:CALCulate :MARKer[111213i4 :X :POSition :STARt <param>