Agilent Technologies ESA-E Series Programmer's Manual page 371

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Error Messages
Greater than 1000 : Personality Specific Error Messages
10239
Opt Freq Ref setting does not match externa l
reference .
This message is generated if Source is set to Externa l
on the properties form under the front-panel Mod e
Setup key and the frequency on the same form is set t o
a frequency that does not match the frequency of th e
signal being used as the external reference .
10232
RF
Signal not found .
This message is generated if there is no signal at th e
center frequency that is greater than 10 dB above th e
displayed average noise level.
10241
RF
Board could not detect any bursts i n
signal .
This message is generated when the trigger is set to R F
Burst and (Option B7E) cannot detect a burst.
10237
RF
Board
LO
Unlocked . Contact service center .
This message occurs if the local oscillator on th e
(Option B7E) is in an unlocked state. This indicates
broken hardware.
10240
RF
Board
RF Osc
Unlocked . Contact servic e
center .
This message occurs if the reference oscillator on th e
(Option B7E) is in an unlocked state . This indicate s
broken hardware.
10238
RF
Board SR
Osc
Unlocked . Contact servic e
center .
This message occurs if the sample rate oscillator on
Option B7E is in an unlocked state. This indicates
broken hardware .
10162
Resolution BW<300kHz .
This error message is a warning that the resolution
bandwidth has been set below 300 kHz . The test results
will not meet GSM specifications.
Chapter 7
7-43

Advertisement

Table of Contents
loading

Table of Contents