Agilent Technologies ESA-E Series Programmer's Manual page 262

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
[:SENSe] :FREOuency Subsectio n
ESA E4404B: 0 Hz, 100 Hz to 6 .78 GH z
ESA E4405B: 0 Hz, 100 Hz to 13 .3 GHz
ESA E4407B, E4408B : 0 Hz, 100 Hz to 27 .0 GHz
Default Unit: Hz
Front Panel
Access :
SPAN/X Scale, Spa n
SPAN/X Scale, Zero Span
Full Frequency Span
[ :SENSe] :FREQuency :SPAN :FULL
Set the frequency span to full scale .
Factory Preset
and *RST:
ESA E4401B, E4411B : 1 .5 GH z
ESA E4402B, E4403B : 3.0 GHz
ESA E4404B: 6.7 GHz
ESA E4405B: 13.2 GHz
ESA E4407B, E4408B : 26 .5 GHz
Front Panel
Access :
SPAN/X Scale, Full Spa n
Previous Frequency Span
[ :SENSe] :FREQuency :SPAN :PREViou s
Set the frequency span to the previous span setting.
Front Panel
Access :
SPAN/X Scale, Last Spa n
Start Frequency
[ :SENSe] :FREQuency :STARt <freq>
[ :SENSe] :FREQuency :STARt ?
Set the start frequency.
Factory Preset
and *RST :
0 Hz
Range : ESA E4401B, E4411B : -80 MHz to 1 .58 GH z
ESA E4402B, E4403B : -80 MHz to 3 .10 GHz
ESA E4404B : -80 MHz to 6 .78 GH z
5-102