Programming Examples
Reading Trace Data Using 32-bit Real Format (GPIB )
Reading Trace Data Using 32-bit
Real Format (GPIB )
/************************************************************ /
/* Reading Trace Data using 32-bit Real Format (HP-IB)
* /
/*
* /
/* This C programming example does the following .
* /
/*
The SCPI instrument commands used are given as
* /
/* reference .
* /
/*
* /
/* -
Opens an HP-IB session at address 18
* /
/* - Clears the Analyzer
* /
/* - Resets the Analyzer
* /
/*
*RS T
/* - Set the input port to the 50 MHz amplitude referenc e
/*
CAL :SOUR :STAT ON
* /
/* - Query for the number of sweep points (for firmware
* /
/*
revisions A .04 .00 and later) . Default is 401 .
* /
/*
SENS :SWE :POIN?
* /
/* - Calculate the number of bytes in the header
* /
/* - Set the analyzer to single sweep mode
* /
/*
INIT :CONT 0
* /
/* - Sets the analyzer center frequency and span to 50 MHz
* /
/*
SENS :FREQ :CENT 50 MHZ
* /
/*
SENS :FREQ :SPAN 50 MHZ
* /
/* - Specify 10 dB per division for the amplitude scale in
* /
/*
and dBm Units
* /
/*
DISP :WIND :TRAC :Y :SCAL :PDIV 10 dB
* /
/*
UNIT :POW DBM
* /
/* - Set the analyzer trace data to 32-bit Real
* /
/*
FORM :DATA : REAL,32
* /
/* - Set the binary order to swap
* /
/*
FORM :BORD SWAP
* /
/* - Trigger a sweep
* /
/*
INIT :IMM
* /
/* - Check for operation complete
* /
/*
*OPC?
* /
/* - Calculate the number of bytes in the trace record
* /
/* - Query the trace data
* /
/*
TRAC :DATA? TRACE1
* /
/* - Remove the "," from the ACSII data
* /
/* - Save the trace data to an ASCII file
* /
/* - Close the session
*/
/************************************************************ /
Chapter 3
3-29