Agilent Technologies ESA-E Series Programmer's Manual page 280

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
[:SENSe] :SWEep Subsectio n
NOTE
This command applies to the following models having firmware revisio n
A.05 .00 or later:
E4401B
E4402B
E4404B
E4405B
E4407B
Example :
:SENSe :SWEep :SEGMent :DATA :MERGe
CSPan,150e06,5e6,1e5,1e5,101, .050,110e06, 5
e6,1e5,1e5,100, .05
This example merges two segments into the segmente d
sweep table . The first segment has a center frequency
of 150 MHz and the second segment has a cente r
frequency of 110 MHz.
Range :
The range depends upon the model number, installe d
options, and firmware revision of your instrument ; see
the Specifications Guide for details . The ranges tha t
apply to all instrument models that can be controlle d
with this command are resolution bandwidth, vide o
bandwidth, and number of points.
Resolution Bandwidth : 1 kHz to 5 MHz
Video Bandwidth : 30 Hz to 3 MH z
Number of Points : 101 to 8192, 2 to 8192 in zero span .
The total number of points for all segments is s 8192.
Default Unit : Fundamental units (without entering the actual units)
Delete All Segmented Sweep Dat a
[ :SENSe] :SWEep :SEGMent :DELet e
Deletes all segmented sweep data .
NOTE
This command applies to the following models having firmware revisio n
A.05.00 or later:
E4401 B
E4402 B
E4404B
E4405B
E4407B
Front Pane l
Access:
Sweep, Segmented, Modify, Delet e
5-120