Agilent Technologies ESA-E Series Programmer's Manual page 139

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Programming Examples
Measuring Harmonic Distortion (RS-232 )
/*
CALC :MARK :TRCK :STAT OF F
/* - Reset VISA timeout to 3 seconds
* /
/* - Set units to dBm
* /
/*
UNIT :POW DBM
* /
/* - Take a sweep
* /
/*
INIT :IMM
* /
/* - Check for operation complete
* /
/*
*OPC?
* /
/* -
Perform the peak search
* /
/*
CALC :MARK :MAX
* /
/* - Read the marker amplitude,this is the fundamental Level* /
/*
CALC :MARK :Y?
* /
/* -
Change the amplitude units to volts
* /
/*
UNIT :POW V
* /
/* - Take a sweep
* /
/*
INIT :IMM
* /
/* - Check for operation complete
* /
/*
*OPC?
*/
/* -
Read the marker amplitude in volts, this is the
* /
/*
fundamental amplitude in volts .
* /
/*
CALC :MARK :Y?
* /
/* -
Read the marker frequency
* /
/*
CALC :MARK :X?
* /
/* -
Measure each harmonic amplitude as follows :
*/
/*
Set the span to 20 MHz
* /
/*
SENS :FREQ :SPAN 20 MHZ
* /
/*
Set the center frequency to the desired harmonic
* /
/*
SENS :FREQ :CENT <freq>
* /
/*
Take a sweep and wait for operation complete
* /
/*
INIT :IMM
* /
/*
*OPC?
*/
/*
Perform peak search
* /
/*
CALC :MARK :MAX
* /
/*
Set VISA timeout to 60 seconds
* /
/*
Activate signal track
* /
/*
CALC :MARK :TRCK :STAT ON
* /
/*
Zoom down to a 100 kHz span
* /
/*
SENS :FREQ :SPAN 10E4
* /
/*
Take a sweep and wait for operation complete
* /
/*
INIT :IMM
* /
/*
*OPC?
*/
/*
Signal track off
*/
/*
CALC :MARK :TRCK :STAT OFF
* /
/*
Reset
VISA timeout to 3 seconds
* /
/*
Perform Peak Search
* /
/*
CALC :MARK :MAX
*/
Chapter 3
3-79

Advertisement

Table of Contents
loading

Table of Contents