Agilent Technologies ESA-E Series Programmer's Manual page 281

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
[:SENSe] :SWEep Subsectio n
Delete Some Segmented Sweep Dat a
[ :SENSe] :SWEep :SEQ(ent :DELete :ROW <integer>
Deletes only the specified segmented sweep data (row) .
NOTE
This command applies to the following models having firmware revisio n
A.05.00 or later:
E4401B
E4402B
E4404B
E4405B
E4407B
Range:
1 to maximum number of segement s
Front Panel
Access :
Sweep, Segmented, Modify, Edit, Delete Segmen t
NOTE
The front panel keys shown for this command deletes the active
segment only. The remote command deletes any specified segment .
Turn On/Off Segmented Sweep
[ :SENSe] :SWEep :SEGMent[ :STATe] OFFIONI01 1
[ :SENSe] :SWEep :SEGMent[ :STATe] ?
Turns on or off segmented sweep . Segmented sweep is a mode where
only bands of interest are measured with the data presented in a singl e
trace. This improves measurement speed and provides more detail an d
resolution in those bands of interest.
NOTE
When segmented sweep is turned on or off, all trace data (includin g
stored trace data) is deleted .
NOTE
This command applies to the following models having firmware revisio n
A.05 .00 or later:
E4401 B
E4402B
E4404B
E4405B
E4407B
Factory Prese t
and *RST :
Of f
Remarks:
Segmented sweep will be turned off if any of th e
following actions occur remotely or manually :
Chapter 5
5-121

Advertisement

Table of Contents
loading

Table of Contents