Language Reference
[:SENSe] :POWer Subsectio n
[ :SENSe] :POWer Subsection
Input Attenuation
[ :SENSe] :POWer[ :RF] :ATTenuation <rel ampl>
[ :SENSe] :POWer[ :RF] :ATTenuation ?
Set the input attenuator. This value is set at its auto value if input
attenuation is set to auto .
Factory Preset
and *RST:
10 dB
Range :
ESA E4401B, E4411B : 0 to 60 dB
ESA E4402B, E4403B, E4404B, E4405B : 0 to 75 dB
ESA E4407B, E4408B : 0 to 65 dB
Default Unit: dB
Front Panel
Access :
AMPUTUDE/Y Scale, Attenuation Auto Ma n
Input Port Attenuator Aut o
[ :SENSe] :POWer[ :Fur
l
:ATTenuation :AUTO OFF1 ON1 01 1
[ :SENSe] :POWer[ :RP] :ATTenuation :AUTO?
Select the input port attenuator range to be set either automatically o r
manually.
On - Input attenuation is automatically set as determined by th e
Reference Level Setting.
Off - Input attenuation is manually se t
Factory Prese t
and *RST :
On
Range :
ESA E4401B, E4411B : 0 to 60 dB
ESA E4402B, E4403B, E4404B, E4405B : 0 to 75 dB
ESA E4407B, E4408B : 0 to 65 dB
Front Panel
Access :
Input/Output (or Input), Input Atte n