Language Reference
CALCulate:LLINe Subsection
Merge Additional Values into the Existing Limit Line
:CALCulate :LLINe[1]~2 :DATA :MERGe
<x-axis>,<ampl>,<connected>{,<x-axis>,<ampl>,<connected> }
Adds the points with the specified values to the current limit line ,
allowing you to merge limit line data . Up to two amplitude values are
allowed for each x value. If too much data is merged, as many points as
possible are merged into the existing limit and then an error i s
reported. Up to 200 points total may be defined for each limit .
• <x-axis> can be frequency or time values as specified b y
: CALCulate :LLINe : CONTrol : DOMain . Frequencies are always i n
Hz . Time is always in seconds . No unit is allowed in this parameter.
• <ampl> amplitude values are always in units of dBm . No unit is
allowed in this parameter .
• <connected> connected values are either 0 or 1 . A 1 means this point
should be connected to the previously defined point to define the
limit line. A 0 means that it is a point of discontinuity and is not
connected to the preceding point . The "connected" value is ignored
for the first point.
Range :
<x-axis> -30 Gs to +30 Gs for time limit s
<x-axis> -30 GHz to +350 GHz for frequency limit s
<ampl> -120 dBm to +100 dBm
<connected> 0 or 1
Front Panel
Access :
Display, Limits, Properties, X Axis Units Freq Tim e
Delete Limit Line
:CALCulate :LLINe[1]12 :DELete
Deletes the selected limit line .
Display, Limits, Modify, Limit On Of f
Display the Limit Line
:CALCulate :LLINe[1]12 :DISPlay OFFIONI01 1
:CALCulate :LLINe[l]I2 :DISPlay?
Controls the display of the current limit line.
Factory Preset
and *RST:
Of f
Front Panel
Access:
5-22
Chapter
5