Error Messages
201 to 799: Device-Specific Errors
729
In <filename> : error reading fil e
Appears when loading data from a limit line or
corrections disk file and a failure to the file occurs .
730
In <filename>, line <numeric_value> : line to o
long
When loading data from a limit line or corrections dis k
file, this message will appear if the length of any line i n
the file exceeds 255 characters .
731
In <command> : bad data coun t
(<numeric_value>) : expected multiple o f
<numeric_value>
This message indicates that the data sent to a
corrections or limit table via the DATA or
MERG E
commands does not have the expected length for the
table. For example, this message would appear if a n
attempt were made to merge 7 numeric values into a
limit table, since each logical entry requires 3 value s
(frequency, amplitude, and connected) .
732
In <filename>, line <numeric_value> : error
parsing tokens
This message may appear when loading data from a
limit line or corrections disk file . It indicates a proble m
in the attempt to break a string of text into tokens.
There may be too few tokens in the string. This
typically happens when there are too few numeric
values in the [DATA] section of a limit or corrections file .
733
In <filename>, line <numeric_value> : <xxx> i s
not numeric
This message may appear when loading data from a
limit line or corrections disk file. It indicates that a
non-numeric token <xxx> was found where a numeri c
token was expected.
735
In <filename> : bad amplitude unit <unit>
This message indicates that unit <unit> is no t
recognized or supported .