Agilent Technologies ESA-E Series Programmer's Manual page 251

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
[:SENSe] :CORRection Subsection
[:SENSe] :CORRection Subsection
Perform Amplitude Correctio n
[ :SENSe] :CORRection :CSET :ALL[ :STATe] OFFIONIO1 1
[ :SENSe] :CORRection :CSET :ALL[ :STATe] ?
Turns On or Off the amplitude corrections . When turned On, only the
correction sets that were turned on are enabled . When turned Off, all of
the correction sets are disabled .
Factory Preset
and *RST:
Of f
Remarks:
To turn On or Off an individual correction set, use :
[ :SENSe] :CORRection :CSET[1]121314[ :STATe] .
Front Panel
Access :
AINPLITUDE/Y Scale, Corrections, Corrections On Of f
Set Amplitude Correction Data
[ :SENSe] :CORRection :CSET[1]I21314 :DATA
<freq>,<rel_ampl>{,<freq>,<rel_ampl> }
[ :SENSe] :CORRection :CSET[1]121314 :DATA?
Sets the amplitude correction data . These frequency/amplitud e
corrections will be applied to the displayed data to correct for syste m
losses/gains outside the analyzer. Four different sets of correction dat a
can be stored .
Example :
: CORR : CSET1 : DATA
900E6,0 .3,1 .0E9,0 .35,1 .3E9,0 . 2
Range:
200 points per set
Default Unit: There are no units on the frequency and amplitud e
pairs. They must be entered in hertz (Hz) and decibels
(dB).
Remarks :
CSET number equivalents to front panel acces s
definitions are as follows :
CSET1 is Antenn a
CSET2 is Cabl e
CSET3 is Other
CSET4 is User
Chapter 5