Agilent Technologies ESA-E Series Programmer's Manual page 265

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
[ :SENSe]:HARMonics Subsection
Remarks:
This command sets the number of harmonics to
measure before computing the total harmoni c
distortion . The minimum number is two (th e
fundamental and the second harmonic) .
Front Panel
Access:
Meas Setup, Harmonic s
Set Harmonic Measurement Sweep Mode
[ :SENSe] :HARMonics :SWEeptime :AUTO OFFIONIO1 1
[ :SENSe] :HARMonics :SWEeptime :AUTO ?
Set auto sweep time per harmonic to off or on .
Factory Preset
and *RST:
On
Remarks:
See [ :SENSe] :HARMonics :SWEeptime <time>
Front Panel
Access :
Meas Setup, ST/Harmonic
Set Harmonic Measurement Sweep Tim e
[ :SENSe] :HARMonics :SWEeptime <time >
[ :SENSe] :HARMonics :SWEeptime ?
Set the sweep time used for measuring each harmonic .
Factory Preset
and *RST:
200/resolution bandwidth (RBW)
Range:
10 ms to the maximum sweep time of the analyzer
Default Unit : seconds
Remarks :
When set to 1 (On), the sweep time for measuring each
harmonic will be set to 200/RBW, where RBW is th e
resolution bandwidth setting in Hz . When set to 0 (Off) ,
you can specify any sweep time within the limits of th e
analyzer, with a minimum of 10 ms . This sweep time is
used only for the zero span measuring of the harmonics,
and not during the search for the fundamental . In this
case, the sweep time used will be the sweep time se t
before the measurement began .
Front Panel
Access :
Meas Setup, ST/Harmoni c
Chapter 5
5-105