Agilent Technologies ESA-E Series Programmer's Manual page 304

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
SYSTem Subsyste m
CTS and will stop transmission if the line becomes unasserted .
Factory Prese t
(no *RST) :
The factory default is On. This parameter is persistent ,
which means that it retains the setting previously
selected, even through a power cycle .
Front Pane l
Access :
non e
Serial Port RTS Setup
:SYSTem :COl4tunicate :SERial[1]I2I3I4I5I6I7I8 :CONTrol :RTS
OFFIONIIBFul l
:SYSTem :COEMunicate :SERial[1]12131415161718 :CONTrol :RTS ?
Sets the hardware pacing (hand-shaking) scheme . Many high spee d
asynchronous modems use this line (paired with CTS) as
receive/transmit pacing. This command is available only on analyzer s
having Option 1AX. Only one Option 1AX can be installed in an
instrument .
Off - indicates that the RTS line should always be asserted
On - indicates that the RTS line should always be unasserted
IBFu1I - selects the input buffer full mode for the RTS line . IBFull
sets the RTS line to indicate when the device is ready to receive .
When the number of received bytes in the input buffer of the devic e
reaches the stop threshold, the device will unassert the RTS line .
When the number of bytes has been reduced to the start threshold ,
the device will assert RTS indicating that it can receive input again .
RTS is sometimes called RFR (ready for receiving) . The device will
also monitor the state of CTS and will stop transmission if that line
becomes unasserted .
Factory Preset
(no *RST) :
The factory default is IBFull . This parameter is
persistent, which means that it retains the settin g
previously selected, even through a power cycle .
Front Panel
Access :
non e
Serial Port Baud Rate Setup
:SYSTem :CONMunicate :SERial[1]I2131415161718[ :RECeive] :BAUD
<baud_rate>
:SYSTem : CONMunicate :SERial[1]I2131415161718[ :RECeive] :BAUD
5-144
Chapters

Advertisement

Table of Contents
loading

Table of Contents