Language Reference
[ :SENSe] :SWEep Subsectio n
Time Gate Length
[ :SENSe] :SWEep :TIME :GATE :LENGth <time >
[ :SENSe] :SWEep :TIME :GATE :LENGth?
Specifies the gate time length in seconds ; for EDGE triggering only. Thi s
command is available only on analyzers having Option 1D6.
Factory Prese t
and *RST:
1µs
Range :
0.3
µs
to 429 seconds
Default Unit: seconds
Front Panel
Access :
Sweep, Gate Setup, Edge Setup, Gate Lengt h
Time Gate Level
[ :SENSe] :SWEep :TIME :GATE :LEVel HIGHILOW
[ :SENSe] :SWEep :TIME :GATE :LEVe1 ?
Selects the level of the gate signal ; this command is for LEVel
triggering only. This command is available only on analyzers having
Option 1D6 .
Factory Preset
and *RST :
High
Front Panel
Access :
Sweep, Gate Setup, Level Setu p
Time Gate Polarity
[ :SENSe] :SWEep :TIME :GATE :POLarity NEGativelPOSitive
[ :SENSe] :SWEep :TIME :GATE :POLarity ?
Selects the polarity of the gate signal ; this command is for EDGE
triggering only. This command is available only on analyzers having
Option 1D6 .
Factory Preset
and *RST:
Positive
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Edge Pos Ne g