To reset the calibration security code to its factory default
N O T E
U1271A/U1272A Service Guide
Figure 1-6 SECUr display
3 Set your new calibration security code.
Press
or
to the left.
Dual
Press
or
Exit
Setup
4 Press
to save the new calibration security code.
5 If the new calibration security code has been successfully
stored, the secondary display will show
down your new calibration security code and store it in a
safe location.
If you have forgotten the correct calibration security code,
you may follow the steps below to reset the calibration
security code to the factory default code (1234).
If you do not have a record (or have lost the record) of the security code,
first try the factory default code, "1234" from the front panel.
Calibration Procedures
Unsecuring the Instrument for Calibration
to move the cursor to the right or
to increment or decrement the digit.
1
. Record
21