Language Reference
IEEE Common Commands
Front Panel
Access:

Reset

*RST
This command presets the instrument to a factory defined condition
that is appropriate for remote programming operation. *RST is
equivalent to performing the two commands :SYSTem:PRESet and
*CLS. This command always performs a factory preset.
The preset performed by *RST is always a factory preset. That is, the
NOTE
same preset performed by :SYSTem:PRESet when
:SYSTem:PRESet:TYPE is set to FACTory.
Front Panel
Access:

Save

*SAV <register>
This command saves the instrument state to the specified instrument
memory register.
Range:
Remarks:
Front Panel
Access:

Service Request Enable

*SRE <integer>
*SRE?
This command sets the value of the service request enable register.
The query returns the value of the register.
Range:

Read Status Byte Query

*STB?
Returns the value of the status byte register without erasing its
contents.
Remarks:
5-8
File, Recall State
Preset
Registers are an integer, 0 to 127
See also commands :MMEMory:LOAD:STATe and
:MMEMory:STORe:STATe
File, Save State
Integer, 0 to 255
See *CLS